How to use the Micro Spectrophotometer more efficiently
The sample size required by the Micro Spectrophotometer is only 0.5-2μ L. No colorimetric cup is required. The sample is dropped directly onto the testing platform with a pipette. The sample automatically forms a liquid column during the measurement process. Absorb paper to avoid cross contamination caused by improper cleaning of the colorimetric cup; It usually have more than one light path (select) automatically by the motor control, and the traditional spectrometer photometric light path to 10 mm, the sample without dilution spectrophotometer, measuring range can be up to 50 times the regular spectrophotometer xenon flash lamp source, rather than a deuterium lamp (uv) and tungsten lamp (visible light), long service life; No preheating, can be detected at any time, detection time is short; When the absorbance value is displayed, the program directly gives the concentration value (nucleic acid, protein and fluorescent dye); It takes up less space in the lab than a traditional spectrophotometer.
Micro Spectrophotometer is an instrument for quantitative and qualitative analysis of substances using spectrophotometry. It is usually used for nucleic acid quantitative analysis, protein quantitative analysis and bacterial growth concentration, and is widely used in proteomics and genomics fields in life science laboratories.
A, the instrument should be placed at 5? In a drying room of 35℃, the relative humidity shall not exceed 85%.
B, the instrument should be placed on a firm and stable workbench, and avoid violent or continuous vibration.
C, indoor lighting should not be too strong, should avoid direct sunlight.
D. The fan should not be blown directly towards the instrument to prevent the light source lamp from affecting the normal use of the instrument due to the unstable luminescence.
E. Stay as far away as possible from high intensity magnetic fields, electric fields and electrical equipment that will generate high frequency waves.
F, the power supply